Clinical Comparison of the accuracy of root canal length determination in permanent teeth by means of Root ZX electronic apex locator, tactile perception and conventional radiography

Document Type : Original Article

Authors

1 Department of Endodontics, Faculty of Dentistry, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.

2 Department of Oral and Maxillofacial Radiology, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.

3 Dentist.

4 Department of periodontics, Faculty of Dentistry, Tehran Islamic Azad University,Tehran, Iran

5 Department of Peadiatrics, Faculty of Dentistry, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.

Abstract

Background and Objectives: Today, several methods have been proposed to determine the root canal length.The purpose of this clinical study was to compare the accuracy of root canal length determination in permanent teeth with ROOT ZX electronic apex locator, tactile perception and conventional radiography. 
Subjects and Methods: Sixty four canals from twenty  molar teeth  from healthy volunteers selected. Access cavities were prepared and working length were determined by means of tactile perception, Root ZX electronic apex locator and conventional radiography and compared with real working length. Data were analyzed via Paired samples T-test.
Results: The accuracy of determination of root canal length by Root ZX, radiography and tactile perception were 93.7, 76.57 and 50% respectively. T test analysis revealed no significant difference between working length determination by Root ZX and real working length (p=0.193). However, there were significant differences between working length determination by   radiographic length and tactile perception with real working length (P<0.001).
Conclusion: The Root ZX apex locator is useful in detecting the root canal length of permanent teeth with radiographic method. However,tactile perception  is not reliable in detecting root canal length

Keywords


1-Walton RE, Torabinejad M. Endodontics : principles and practice . 4th ed. St. Louis: Saunders Elsevier; 2009. P.  230-58.   
2-Huang L. An experimental study of the principle of electronic root canal measurement. J Endod 1987;13(2):60-4.
3-Inoue N. An audiometric method for determining the length of root canals. J Can Dent Assoc 1973;39(9):630-6.
4-Goldman M, Pearson AH, Darzenta N. Endodontic success--who's reading the radiograph? Oral Surg Oral Med Oral Pathol 1972;33(3):432-7.
5-Ricard O, Roux D, Bourdeau L, Woda A. Clinical evaluation of the accuracy of the Evident RCM Mark II Apex Locator. J Endod 1991;17(11):567-9.
6-Ingle J, Bakland LK, Baurngartner JC. Ingle's endodontics. 6th ed. Hamilton, Ontario: BC Decker; 2008. P. 516-24.
7-Seidberg BH, Alibrandi BV, Fine H, Logue B. Clinical investigation of measuring working lengths of root canals with an electronic device and with digital-tactile sense. J Am Dent Assoc 1975;90(2):379-87.
8-Subramaniam P, Konde S, Mandanna DK. An in vitro comparison of root canal measurement in primary teeth. J Indian Soc Pedod Prev Dent 2005;23(3):124-5.
9-Frank AL, Torabinejad M. An in vivo evaluation of Endex electronic apex locator. J Endod 1993;19(4):177-9.
10- McDonald NJ. The electronic determination of working length. Dent Clin North Am.1992 Apr ;36(2):293-307.
11-McDonald NJ, Hovland EJ. An evaluation of the Apex Locator Endocater. J Endod 1990;16(1):5-8.
12-Schilder H. Filling root canals in three dimensions. Dent Clin North Am 1967:723-44.
13-Shahrabi M, Seraj B, Nekoofar MH, Moshrefian S, Kharazi Fard MJ. An evaluation on the accuracy of an electronic apex Locator (EAL) in the determination of working length inprimary teeth (In-vitro). J Dent Med 2004;17(1):32-40.
14-Zokaei S, Fallahatnejad M, Parirokh M, Eftekhar vaghefi H, Mohammad Alizadeh S. [Accuracy of electronic apex locator (EAL) and radiographic film (RF) and digital radiography (DR) to determine working length]. J Dent School Shahid Beheshti Univ Med Sci 2010;28(2):71-9. [In Persian]
15-Mokhtari F, Mahabadi AM, Ghafarian N, Ghate A, Razavi A. Comparison of the accuracy of Root ZX and Novapex with radiography: an in vivo study. J Dent Med 2012;25(1):19-26.
16-Singh SV, Nikhil V, Singh AV, Yadav S. An in vivo comparative evaluation to determine the accuracy of working length between radiographic and electronic apex locators. Indian J Dent Res 2012;23(3):359-62.
17-Chakravarthy Pishipati KV. An In Vitro Comparison of Propex II Apex Locator to Standard Radiographic Method. Iran Endod J 2013;8(3):114–7.
18- Alidadi A. Evaluation of Reypex electronic apex finder accuracy in determining root canal length of permanent teeth [dissertation]. Ahvaz: Joundi Shapour University of Medical Science; 2004. P. 50-4. [In Persian]
19-Kishor KM. Comparison of working length determination using apex locator, conventional radiography and radiovisiography: an in vitro study. J Contemp Dent Pract 2012;13(4):550–3.
20-Parekh V, Taluja C. Comparative study of periapical radiographic techniques with apex locator for endodontic working length estimation: an ex vivo study. J Contemp Dent Pract 2014;12(2):131-4.
21-Kqiku L, Städtler P. Radiographic versus electronic root canal working length determination. Indian J Dent Res  2011;22(6):777-80.
22-Shahi S, Rahimi S, Salem Milani A, Asghari Jafarabadi M, Djoorabbaf Naghsh GR. Comparison Between the Accuracy of NovApex Apex Locator and Radiographs in Determining Radiographic Apex. Iran Endod J 2011;6(2):65-8. 
23-Ravanshad S, Adl A, Anvar J. Effect of working length measurement by electronic apex locator or radiography on the adequacy of final working length: a randomized clinical trial. J Endod 2010;36(11):1753-6.