Comparison of working length determination by NSK electronic apex locator with the radiographic method on necrotic deciduous molars: In vitro

Document Type : Original Article

Authors

1 Department of Dental Pediatrics, Dental School, Ahvaz Jundishapur Medical Science, Ahvaz, Iran.

2 Department of Oral and Maxillofacial Radiology, Dental School, Ahvaz Jundishapur Medical Science, Ahvaz, Iran.

3 Department of Orthodoncy, Dental school, Ahvaz Jundishapur Medical science, Ahvaz, Iran. 4-Private Practice.

4 Private Practice.

5 Dental student. Dental school, Ahvaz Jundishapur Medical Science, Ahvaz, Iran.

Abstract

Background and Objective: Working length determination is a basic procedure in endodontic treatment of permanent and deciduous teeth. Radiography is the conventional method of canal length determination. In this method the distance between tip of the file inserted in the root canal and the tip of the radiographic apex is measured. The purpose of this study was to compare working length determination by NSK (NSK; Japan) electronic apex locator (EAL) with the radiographic method on necrotic deciduous molars.
Subjects and Methods: This in vivo study performed on 60 root canals of 20 primary molars. Canal length measured by NSK (NSK; Japan) electronic apex locator compared with radiographic method. Paired samples test was used for the statistical analysis.
Results: The mean canal length of D teeth was 13.82 mm with the radiologic method, and 13.83 mm with EAL. There were no significant differences between the two methods (P= 0.81). The average canal length of  E  teeth  was 15.03 mm with the radiologic method, and 15 mm with EAL showed no statistically significant differences between the two methods (P= 0.57).Totally the canal length was 14.43mm with the radiologic method, and 14.42 mm with EAL with no statistical significant differences (P=0.85).
Conclusion: According to these results of this study we can use EAL instead of radiographic method in working length determination of primary teeth.

Keywords


1-Mc Donald RE, Avery DR, Dean JE. Dentistry for the child and Adolescent. 8th ed. New York: Mosby; 2006. p.595.
2-Shahrabi M, Seraj B, Heidari A. [Invivo evaluation of the accuracy of an electronic apex locator in root canal length determination in primary teeth]. J of Dent Med 2006;19(1):79-83.
3-Kim E, Marmo M, Lee CY, Oh NS, Kim IK. An in vivo comparison of working length determination by only root-Zx apex locator versus combining root-Zx apex locator with radiographs using a new impression technique. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008;105(4):e79-83.
4-Krajczár K, Marada G, Gyulai G, Tóth V. Comparison of radiographic and electronical working length determination on palatal and mesio-buccal root canals of extracted upper molars. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008;106(2):e90-3.
5-Sadeghi SH, Abolghasemi M. A comparison between the ray pex 5 apex locator and conventional radiography for determining working length of straight and curved canals. IEJ 2007;2(3):101-5.
 6-Gordon MP, Chandler NP. Electronic apex locators. Int Endod J 2004;37(7):425-37.
7-Akisue E, Gavini G, de Figueiredo JA. Influence of pulp vitality on length determination by using the Elements Diagnostic Unit And Apex Locator. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007;104(4):e129-32.
8-Nekoofar MH, Ghandi MM, Hayes SJ, Dummer PM. The fundamental operating principles of electronic root canal length measurement devices. Int Endod J 2006;39(8):595-609.
9-Martínez‐Lozano M, Forner‐Navarro L, Sánchez‐Cortés JL, Llena‐Puy C. Methodological considerations in the determination of working length. Int Endod J 2001;34(5):371-6.
10-Seltzer S, Endodontdogy-biologic considerations in endodontic procedures. 2nd ed,1988:24-24.
11-Witte RA. Electronic Test Instruments: Analog and Digital Measurements. 2nd ed. New Jersey: Prentice Hall; 2002.
12-Stavrianos C, Vasiliadis L, Stavrianou I, Louloudiadis A. In vitro evaluation of the ability of Ray-Pex 5 to determine the working length in teeth with simulated apical root resorption. Balkan J Stomatol 2009;13(2):96-8.
13-Kaufman AY, Keila S, Yoshpe M. Accuracy of a new apex locator: an in vitro study. Int Endod J 2002;35(2):186-92.
14-Bernardes RA, Duarte MA, Vasconcelos BC, Moraes IG, Bernardineli N, Garcia RB, et al. Evaluation of precision of length determination with 3 electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator, and RomiAPEX D-30. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007;104(4):e91-4.
15-Lee SJ, Nam KC, Kim YJ, Kim DW. Clinical accuracy of new apex locator with an automatic compensation circuit. J Endod 2002;28(10):706-9.
16-Herrera M,  Ábalos C, Planas AJ, Llamas R. Influence of apical constriction diameter on Root ZX apex locator precision. J Endod 2007;33(8):995-8.
17-Torabinejad M, Walton ER. [Endodontics principles and practice]. Tehran: Teimorzadeh; 2007. P. 20-35. [In Persian]
18-Bodur H, Odabaş M, Tulunoğlu O, Tinaz AC. Accuracy of two different apex locators in primary teeth with and without root resorption. Clin Oral Investing 2008;12(2):137-41.
19-Martinez-Lozano MA, Forner-Navarro L, Sanchez-Cortes JL, Llena-Puy C. Methodological considerations in the determinationof working length. Int Endod. 2001;  J 34: 371–376.
20-Lucena-Martín C, Robles-Gijón V, Ferrer-Luque CM, de Mondelo JM. In vitro evaluation of the accuracy of three electronic apex locators. J Endod 2004;30(4):231-3.
21-Kaufman AY, Keila S, Yoshpe M. Accuracy of a new apex locator: an in vitro study. Int End J 2002;35(2):186-92.
22-Javidi M, Moradi S, Rashed R, Raziee L. In vitro comparative study of conventional radiography and Root ZX apex locator in determining root canal working length. N Y State Dent J 2009;75(4):48-51.
23-Hoer D, Attin T. The accuracy of electronic working length determination. Int Endod J 2004;37(2):125-31.
24-Subramaniam P, Konde S, Mandanna DK. An in Vitro comparison of root canal measurement in primary teeth. J Indian Soc pedod Prev Dent 2005;23(3):124-5.
25-Vieyra JP, Acosta J, Mondaca JM. Comparison of working length determination with radiographs and two electronic apex locators. Int Endod J 2010;43(1):16-20.
26-Andrian  S, Iovan G, Georgescu  A, Arnăuţeanu C, Stoleriu S. A comparative study on the precision of the radiological and electronic methods for determining root canal working lengths. Int J Med Dent 2012;2(3):172-77.