Derivation of Equivalent Square Field from Antisymetric Field Mega Voltage Photon Beams by Taylor Expantion

Document Type : Original Article

Authors

1 Department of Physics, Ahvaz Branch, Islamic Azad University, Ahvaz, Iran

2 Department of Medical Physics Ahvaz, JundishapurUniversity of Medical Sciences, Ahvaz, Iran.

3 Department of Medical Physics, Isfehan University of Medical Sciences, Iran.

Abstract

Background and Objective: Determining the symmetric fields equivalent to antisymetric fields ones is an important task in radiotherapy .There are many semiempirical methods based on  exprimatal data.In this study, attempts have been made to give a formula on the basic of logical and physics principles r.
Subjects and methods: Percentage of depth dose is a function of dimensions of the radiation field (x,y) in constant depth. This means that PDD=f (x,y). It have been changed the variables  and expand PDD in Taylor’s series around the refrence field (10×10). It was ignored degrees higher than two. The coefficient of expansion can be derived from the dosimetry by four experimental measurments. By applying this formula, the PDD for any radiation field is ditermineded.
Results: The calculated PDD data of radiation fields larger than than 5×5have been compared with the experimental data of radiation energies 6 and 18MV.
Conclusion: The difference between the calculated and experimental data of the small radiation fields is more than that of the large fields because dimension of ionization chamber is larger in comparison with the field size. This error is not important since the usual field sizes in radiotherapy are not small. Therefore, this formula is applicable in radiotherapy.
Sci Med J 2012;10(6):675-683

Keywords


 
1-Clarkson JR. A note on depth doses in fields of irregular shape. Br J Radiol. 1941; 14: 265-8.
2-Faiz MK. The Physics of radiation therapy. 1994; P.219.
3-Sterling TD, Perry H, Katz L. Automation of radiation treatment planning. IV. Derivation of a mathematical expression for the per cent depth dose surface of cobalt-60 beams and visualisation of multiple field dose distribution. Br J Radiol. 1964; 37: 544-50.
4-Wrede TF, Edwards MSC. An intercomparison between two methods of obtaining percentage depth doses for irregular shaped field and comparison of each method with experimental data for co-60 and 10MV X rays. Br J Radiol. 1979; 52(39): 40-8.
5-Angelo F, Monti AO, Milena F, Stefania G. Anequivalent square method for irregular photon field. Medical Dosimetry. 1995; 20(4): 275-7.
6-Tahmasebi MJ, Ghorbani M, Mehdizadeh A. New formula for calculation of cobalt60 percent depth dose. Biomedizinische Tecknik. 2005; 50 (1): 5-6.
7-Tahmasebi MJ, Behrooz MA, Shahbazian H, Hamzian N, Karbalaee M. Determination of the equivalent field size of asymmetric fields in photon beam therapy analytically based on physical theories. Scientific Medical Journal. 2009; 8(1): 38-55.