Investigating the Effect of Magnetic Field on Radiation Dose Distribution in Radiotherapy using Photon and Electron Beams of Linear Therapeutic Accelerators

Document Type : Original Article

Authors

1 Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.

2 Cancer Research Center, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.

3 Department of Clinical Oncology, School of Medicine, Golestan Hospital, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.

4 Department of Radiology Technology, School of Allied Medical Sciences, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.

5 Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran

10.22118/jsmj.2023.426038.3297

Abstract

Background and Objectives Magnetic fields can be used in radiation therapy to reduce electron contamination and improve dose delivery accuracy. MRIgRT systems use magnetic fields to track the position of the tumor during treatment and precisely deliver the dose from electron beams to the tumor, which will lead to improved treatment outcomes and reduced side effects. Subjects and Methods The MCNP 6.1 Monte Carlo code was used to simulate the Varian 2100 C/D LINAC in both photon and electron modes. Percentage depth dose curves, dose profiles, and the fluence of contaminating electrons and photons were calculated. Dose profile penumbra and dose differences were calculated for different modes. In the second phase of the study, a constant 1.5 Tesla longitudinal magnetic field was applied to a water phantom that was aligned with the direction of the radiation beam. Results The MD reduced the surface dose by 8.3% and the dose profile penumbra by 5.6% at the surface of the water phantom. The MD removes all contaminating electrons from the radiation field without affecting the number of photons. The application of a 1.5 Tesla longitudinal magnetic field increased the dose by 4% in the maximum dose depth region and reduced the penumbra by 20% and the off-axis dose by 57% at the same depth. Conclusion The MD reduces surface dose, off-axis dose, and dose profile penumbra. The longitudinal magnetic field reduces penumbra and off-axis dose in electron beams.

Highlights

Morteza Hashemizadeh [PumMed] [Google Scholar]

Mansour Zabihzadeh [PumMed] [Google Scholar]

 Hojatollah Shahbazian [PumMed] [Google Scholar]

 Jafar Fatahi-Asl [PumMed] [Google Scholar]

 Marziyeh Reshadatian [PumMed] [Google Scholar]

Keywords


[1] Semwal M. Khan's the physics of radiation therapy. 6th ed. Wolters Kluwer (India) Pvt. Ltd., New Delhi; 2020.P.134. [DOI: 10.4103/jmp.JMP_17_20] [PMCID] [2] Kueng R, Oborn BM, Roberts NF, Causer T, Stampanoni MFM, Manser P, et al. Towards MR-guided electron therapy: Measurement and simulation of clinical electron beams in magnetic fields. Physica Medica. 2020;78:83-92. [DOI: 10.1016/j.ejmp.2020.09.001] [PMID] [3] Parsai EI, Shvydka D, Pearson D, Gopalakrishnan M, Feldmeier JJ. Surface and build-up region dose analysis for clinical radiotherapy photon beams. Applied Radiation and Isotopes. 2008;66(10):1438-42. [DOI: 10.1016/j.apradiso.2008.02.089] [PMID] [4] Mesbahi A, Mehnati P, Keshtkar A. A comparative Monte Carlo study on 6MV photon beam characteristics of Varian 21EX and Elekta SL-25 linacs. International Journal of Radiation Research. 2007;5(1):23-30. [Link] [5] Medina AL, Teijeiro A ,Salvador F, Medal D, Vazquez J, Salgado M, et al. Comparison between TG-51 and TRS-398: electron contamination effect on photon beam-quality specification. Physics in Medicine and Biology. 2004;49(1):17-32. [DOI: 10.1088/0031-9155/49/1/002] [PMID] [6] Yani S, Dirgayussa IGE, Rhani MF, Soh RCX, Haryanto F, Arif I. Monte Carlo study on electron contamination and output factors of small field dosimetry in 6 MV photon beam. Smart Science. 2016;4(2):87-94. [DOI: 10.1080/23080477.2016. 1195609] [7] Damrongkijudom N, Oborn B, Butson M, Rosenfeld A. Measurement of magnetic fields produced by a "magnetic deflector" for the removal of electron contamination in radiotherapy. Australasian physical & engineering sciences in medicine. 2006;29(4):321-7. [DOI: 10.1007/BF03178398] [PMID] [8] Zhu TC, Palta JR. Electron contamination in 8 and 18 MV photon beams. Medical physics. 1998;25(1):12-9. [DOI: 10. 1118/1.598169] [PMID] [9] Sjögren R, Karlsson M. Electron contamination in clinical high energy photon beams. Medical physics. 1996;23(11):1873-81.
[DOI: 10.1118/1.597750] [PMID] [10] Yano M, Araki F, Ohno T. Geant4 Monte Carlo investigation of the magnetic field effect on dose distributions in low-density regions in magnetic resonance image-guided radiation therapy. Physica Medica. 2019;68:17-34. [DOI: 10.1016/j.ejmp. 2019.11.005] [PMID] [11] Mutic S, Dempsey JF. The ViewRay System: Magnetic Resonance–Guided and Controlled Radiotherapy. Seminars in Radiation Oncology. 2014;24(3):196-9. [DOI: 10.1016/j.semradonc.2014. 02.008] [PMID] [12] Roberts DA, Sandin C, Vesanen PT, Lee H, Hanson IM, Nill S, et al. Machine QA for the Elekta Unity system: A Report from the Elekta MR-linac consortium. Medical physics. 2021;48(5):e67-e85. [DOI: 10.1002/mp.14764] [PMID] [PMCID] [13] Weinhous MS, Nath R, Schulz RJ. Enhancement of electron beam dose distributions by longitudinal magnetic fields: Monte Carlo simulations and magnet system optimization. Medical physics. 1985;12(5):598-603. [DOI: 10.1118/1.595681] [PMID] [14] Bielajew AF. The effect of strong longitudinal magnetic fields on dose deposition from electron and photon beams. Medical physics. 1993;20(4):1171-9. [DOI: 10.1118/1.597149] [PMID] [15] Litzenberg DW, Fraass BA, McShan DL, O'Donnell TW, Roberts DA, Becchetti FD, et al. An apparatus for applying strong longitudinal magnetic fields to clinical photon and electron beams. Physics in medicine and biology. 2001;46(5):N105-15.
[DOI: 10.1088/0031-9155/46/5/401] [PMID] [16] Chen Y, Bielajew AF, Litzenberg DW, Moran JM, Becchetti FD. Magnetic confinement of electron and photon radiotherapy dose: a Monte Carlo simulation with a nonuniform longitudinal magnetic field. Medical physics. 2005;32(12):3810-8. [DOI: 10.1118/1.2011091] [PMID] [17] Ghila A, Steciw S, Fallone BG, Rathee S. Experimental verification of EGSnrc Monte Carlo calculated depth doses within a realistic parallel magnetic field in a polystyrene phantom. Medical Physics. 2017;44(9):4804-15. [DOI: 10.1002/ mp.12413] [PMID] [18] Goorley T, James M, Booth T, Brown F, Bull J, Cox LJ, et al. Initial MCNP6 Release Overview. Nuclear Technology. 2012;180(3): 298-315. [DOI: 10.13182/NT11-135] [19] Goorley T. MCNP6. 1.1-beta release notes. LA-UR-14-24680; 2014. [Link] [20] Bull J. Magnetic Field Tracking Features in MCNP6. LA-UR-11-00872, Los Alamos National Laboratory; 2011. [Link] [21] Kelner SR, Prosekin AY, Aharonian FA. Synchro-curvature radiation of charged particles in the strong curved magnetic fields. The Astronomical Journal. 2015;149(1):33. [DOI: 10. 1088/0004-6256/149/1/33] [22] Oborn BM, Gargett MA, Causer TJ, Alnaghy SJ, Hardcastle N, Metcalfe PE, et al. Experimental verification of dose enhancement effects in a lung phantom from inline magnetic fields. Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology. 2017;125(3):433-438. [DOI: 10.1016/j.radonc.2017.09.012] [PMID] [23] Alnaghy SJ, Begg J, Causer T, Alharthi T, Glaubes L, Dong B, et al. Technical Note: Penumbral width trimming in solid lung dose profiles for 0.9 and 1.5 T MRI-Linac prototypes. Medical physics. 2018;45(1):479-87. [DOI: 10.1002/mp.12680] [PMID]